Description
AR-203 Ventilator (Air compressor inside the machine)
Description:
Display mode |
High-definition 10.4” TFT color LCD screen |
Ventilation mode |
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH, IMV, VCV, PCV |
Ventilation parameters |
Tidal volume 20~2000 ml Rate 1~120 bpm SIMV rate 2~20 bpm I:E 4:1~1:8 Pressure trigger sensitivity -1.0-2.0kPa Flow trigger sensitivity 1-20L/min PEEP 0~2.0Kpa Pressure range 3~60cmH2O Pressure support 3~60cmH2O SIGH 1.5 times the inspiratory time (60-120 adjustable) Inspiratory platform 0~50% Oxygen concentration 21~100% Pressure limitation 5~60cmH2O Pressure slope 1~10gear Flow rate setting 1~10gear Expiratory trigger pressure -1.0~2.0Kpa Time control 1~10s |
Parameters for ventilation monitoring |
Tidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, Inspiratory platform (autonomous inspiratory rate, lung compliance, airway resistance, flow -volume loop, pressure-volume loop) |
Monitoring of oxygen concentration |
21%~100% |
Security Alarm System |
|
Oxygen Concentration alarm |
upper limit setting range 21%~100% low limit setting range 10%~80% |
Airway pressure alarm |
upper limit setting range 0.3~6.0kPa low limit setting range 0.2~5.0kPa |
Per-minute ventilation volume alarm |
upper limit setting range 3.0~30L/min low limit setting range 1.0~10L/min |
Respiratory rate |
upper limit setting range 50~99bpm low limit setting range 0~50bpm |
Tidal volume |
upper limit setting range 10~2000ml low limit setting range 0~1800ml |
Sustained high-pressure alarm |
it will give alarm when stress have consistently been higher than 2.5kPa |
Suffocation alarm |
it will give sound and light alarm if there is no tidal volume input for 15 seconds |
Power alarm Gas shortage alarm Intubation off alarm |
|
Power |
AC 220V±10% 50Hz±1Hz UPS, Storage battery |