Nihon Kohden
ASP-1800T Syringe Pump (Dual Channel)
- Weight:
- 11.00 KGS
- Shipping:
- Calculated at Checkout
Description
ASP-1800T Syringe Pump (Dual Channel)
Features:
1. Simplified, Intuitive Touch
- Independent operating channels, and with Online function----AB channels can be worked cyclically and repeatedly;
- 3.5-inch large color display, making clinical use faster and more convenient;
- Independent power switch, conducive to clinical use and energy conservation;
-Tube management: a standard fitted side clamp to prevent infusion tube entanglement and enhance infusion overview and control.
2. Safe, Efficient and Practical
- Comply with IP44 high waterproof rating for all clinical application scenarios;
- DPS dynamic pressure display, multi-pressure alarm level, real-time protection of patient safety;
- Anti-bolus & anti-siphon functions increase patient safety;
- Auto & manual Bolus can be switched; KVO rate is adjustable for more clinical use;
- Change the flow rate during infusion without stooping the infusion;
- Large-scale & wide-angle color alarm lights to protect patients in real time.
3. Intelligent Design, Easy to Use
- 5 infusion modes are available to meet the needs of different departments;
- 360 degree rotatable, vertical and horizontal mounting clamp to meet the needs of the clinical environment;
- More than 2,000 historical records for review and export, which is convenient for users to use;
- Multi-language can be switched and customized to meet various geographical needs.
Technical Specifications
PARAMETER |
SPECIFICATION |
Applicable Syringes |
5ml,10ml,20ml,30ml,50ml,60ml of any standard |
Accuracy |
+/- 2% with correct calibration |
Rate Mode |
50ml(60ml): 0.1ml/h-2000.0ml/h (0.1ml/h or 1ml/h step) 30ml: 0.1ml/h-1200.0ml/h (0.1ml/h or 1ml/h step) 20ml: 0.1ml/h-800.0ml/h (0.1ml/h or 1ml/h step) 10ml: 0.1ml/h-400.0ml/h (0.1ml/h or 1ml/h step) 5ml: 0.1~200ml/h (0.1ml/h or 1ml/h step) (Programmable 0.01ml/h~9999.9ml/h) |
Time Mode |
SOL. VOL: 0.1-9999.9ml (0.1ml step) TIME: 00:01-99:59min (1min step) |
Dose Mode |
Volume: 0.1-999.9ml (0.1ml step) DRUG MASS: 0.1-999.9mg (0.1mg step) DOSE: 0.001-9999ug/kg/min (0.001ug/kg/min step) WEIGHT: 0.1-300kg (0.1kg the least step) Dose unit: ng/kg/min, ug/kg/min, mg/kg/min, g/kg/min, ug/kg/h, mg/kg/h, g/kg/h, ug/min, mg/min, g/min, ug/h, mg/h, g/h, U/min, U/h, IU/h, U/kg/h, ml/h |
Total Volume Range |
0.1-9999.9ml (0.1 ml step) |
Drug library Mode |
10 categories of drugs, 17kinds of dose units. Makes drug dispensing more convenient and safer. |
Sequential Mode |
Available |
Purge Rate |
50ml(60ml): 1500.0ml/h, 30ml: 1200.0ml/h, 20ml: 800.0ml/h 10ml: 400.0ml/h, 5ml: 200ml/h Programmable and adjustable |
KVO Rate |
0.1~5.0 ml/h, Programmable and adjustable |
DPS Dynamic Pressure |
Three levels setting LOW (60(±15)kPa), MEDIUM (90(±15)kPa) or HIGH (120(±15) kPa) |
Alarms |
Syringe Holder Loose, Syringe Plunger Disengagement, Syringe Flange Insertion Error, Syringe Not Calibrated, Occlusion, Syringe Empty, Rate Abnormal, Battery Empty, No Battery, Infusion Complete, Syringe Near Empty, No Action, Battery Low Three levels of alarm volume LOW, MEDIUM or HIGH |
Bolus Rate |
5ml: 0.1~200ml/h,10ml:0.1-400ml/h,20ml:0.1-800ml/h, 30ml: 0.1-1200ml/h,50ml(60ml):0.1-2000ml/h Programmable and adjustable, Automatic/Manual bolus available |
Bolus Volume |
Range: 1.0~50.0ml, (0.1ml step) |
Interface |
RS-232 |
Brightness |
1~10 levels can be selected |
History Records |
2000 events stored can be checked and downloaded |
Power Source |
AC power supply: AC 100 to 240V(+/-10%), 50/60Hz; DC Power supply: 12~15V, power>15W; Built-in Lithium Battery: Rechargeable, 11.1V, 2000mAh. 8 hours of continuous operation |
Accessories |
AC power line, Operational manual |
Protection Class |
Class I, Type CF, IP44 |
Dimension/N.W |
350*150*220mm, 4kg |
Packing size/G.W |
400mm*200mm*2500mm, 4.5KG |
Contact Us Today for more details